You signed in with another tab or window. Reload to refresh your session.You signed out in another tab or window. Reload to refresh your session.You switched accounts on another tab or window. Reload to refresh your session.Dismiss alert
{{ message }}
Repository navigation
Stage. Mapping arbitrary units to phisical units #288
I have a question that rose while testing the ASI stage but I find applies to others.
The move_by method takes an value and it Move axis by given amount. It is not specified the amount in physical units. In the case of the ASI this is translated into 1/10 microns, and that value must be different for any stage.
I cannot find anywhere a mapping of these arbitrary units and the physical units. Where is this to be defined?
I guess, although we would then have to implement some well defined mechanism to signal that there is no physical unit and the user needs to provide one.
Thinking about this I would say that more and more info will be encoded on the stage so we probably should read units from the stage if possible.
Maybe have cockpit do,
try:
get stage units from microscope
except:
try:
get stage units from depot config
except:
fail, no stage units.
The one drawback of this is how do you overwrite the stage units if they are mad. Maybe the opposite is better and try read from config, then the stage if both fail then error.
Activity
iandobbie commentedon Oct 26, 2023
Microscope stages are just defined in terms of "units" whatever the stage wants to use. Any mapping to real units is defined in cockpit, see https://microscope-cockpit.org/depot-config#stages
juliomateoslangerak commentedon Oct 26, 2023
I see. Thanks
Does it make sense to implement some function to get this information from the stage? Best effort
iandobbie commentedon Oct 26, 2023
I guess, although we would then have to implement some well defined mechanism to signal that there is no physical unit and the user needs to provide one.
iandobbie commentedon Nov 8, 2023
Thinking about this I would say that more and more info will be encoded on the stage so we probably should read units from the stage if possible.
Maybe have cockpit do,
try:
get stage units from microscope
except:
try:
get stage units from depot config
except:
fail, no stage units.
The one drawback of this is how do you overwrite the stage units if they are mad. Maybe the opposite is better and try read from config, then the stage if both fail then error.